Guidance and manufacturer's declaration - electromagnetic emissions
The BM1003 is intended for use in the electromagnetic environment specified below. The customer or the user of the
BM1003 should assure that it is used in such an environment.
Emissions
Compliance
Electromagnetic environment – guidance
test
The BM1003 uses RF energy only for its internal function. Therefore, its RF
RF emissions
Group 1
emissions are very low and are not likely to cause any interference in nearby
CISPR 11
electronic equipment.
RF emissions
Class B
CISPR 11
Harmonic
The BM1003 is suitable for use in all establishments other than domestic and
emissions
Not applicable
those directly connected to the public low-voltage power supply network that
IEC 61000-
supplies buildings used for domestic purposes.
3-2
Voltage
fluctuations/
flicker
Not applicable
emissions
IEC 61000-
3-3
90
Guidance and manufacturer's declaration – electromagnetic immunity
The BM1003 is intended for use in the electromagnetic environment specified below. The customer or the user of the
BM1003 should assure that it is used in such an environment.
IEC 60601
Compliance
Immunity test
test level
level
Electrostatic
± 6 kV
discharge
± 6 kV
contact
(ESD)
contact
± 8 kV air
IEC 61000-4-2
Power
frequency
(50/60 Hz)
3 A/m
3 A/m
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
be at least 30%.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.
91