Installation
Selecting a Location
Choose a location for the FBV18FPSA that provides a
clearance of 12" at the top, 2" at the rear and 2" on each
side.
Appropriate electrical power must be available. Attach the
freezer to the facility's electrical supply as directed by the
National Electrical Code.
Installation Instructions
Remove all packing materials, including the bottom
cushions, foam pads, and tape inside the unit. Tear off all
protective fi lm on the doors and on the freezer body.
• Place the unit in a well ventilated place, on fl at sturdy
ground.
• Avoid locating the unit in moist or humid areas. Excess
moisture in the air may cause frost to form quickly on the
evaporator and cabinet, requiring frequent defrosting.
• Locate the unit away from direct sunlight and other heat
sources. Direct sunlight may aff ect the performance of
the unit.
Let the freezer stand upright for approximately 24 hours
before connecting it to power. This reduces the possibility
of a malfunction in the cooling system from improper
handling during transportation.
While moving the unit, DO NOT tilt it beyond 45° from
upright.
Leveling the Unit
It's very important for the freezer to be level. If the freezer is
not leveled during installation, the doors may not close or
seal properly, causing cooling, frost or moisture problems.
To level your freezer, rotate the leveling feet clockwise
(to raise) or counter-clockwise (to lower).
Note: Have someone push against the top of the freezer to
help take some weight off the leveling foot, making it easier
to adjust.
12
Electrical Connection
Determine the total amount of current presently being
used by other apparatus connected to the circuit that will
be used by this freezer. It is critical that this added current
demand and other equipment on this circuit not exceed the
rating of the fuse or circuit breaker in use.
The frequency and nominal voltage requirements for the
unit are specifi ed on the data plate. Supply this unit with an
electrical source that meets these requirements only. Low
line voltage is often the cause of service complaints. With
the unit running, check that the line voltage is within
-6% to +10% of that specifi ed on the data plate.
CAUTION: Be sure the voltage supplied to
the freezer is equal to that specifi ed on the
data plate.
WARNING: For personal safety, this unit
must be properly grounded before use.
General Purpose units must be connected to a grounded
outlet matching the nameplate and / or the information
furnished in this manual. If you are not sure about the
outlet, you should contact a qualifi ed electrician for
assistance.
The Unit should always be connected to a dedicated power
source.
WARNING: DO NOT REMOVE, under any
circumstance, the grounding prongs from
the 3-prong power cord supplied with all
units.
WARNING: DO NOT USE electrical
extension cords, it may result in voltage
loss and possible hazardous operation.