Technical Specifications
Technical Specifications
Fume Hood/
Ventilation System
Location
Ambient Temperature
Humidity
Altitude
Pollution Degree
Electrical Supply
Max Power Consumed
Materials
Weight
Dimensions
IP Classification
RF Type (Sensor)
Operating Frequency
Range
Radar Peak Output
Power
Sensor Cable Length
2
A laboratory fume hood with the capacity
to handle leakage of solvents and provide
suitable airflow to prevent the build-up of
flammable vapors.
Indoor use only. No wet locations.
Biotage
Liquid Level Sensor must be
®
placed on a stable, and flat level surface.
Operating: 15°C to 32°C (59°F to 90°F).
Storage and transportation: -25°C to 60°C
(-13°F to 140°F).
20% to 90% RH (non-condensing).
Operation: up to 2000 m (6562 ft).
The Biotage
Liquid Level Sensor kit shall be
®
installed and used in a level 2 environment
in accordance with EN 60664-1.
24 V
500 mA
PEEK (tube holder), PPS (sensor house,
cap, and universal cap adapter), PP (hub),
TPE partly coated with PO (sensor cable)
Hub: 133.5 g (0.29 lbs)
Sensor: 64.2 g (0.14 lbs)
Hub: 157 x 85 x 32 mm
(6.2" x 3.3" x 1.3")
Sensor: 63 x 55 x 27 mm
(2.5" x 2.2" x 1.1")
IP54 for the sensor
Transceiver
57 to 64 GHz
18 dBm
1.8 m (70.9")