Connect the Liquid Level Sensors to the System
5. Snap the sensor onto the tube holder as shown in Figure 7.
Figure 7. Snap the sensor onto the tube holder.
6. Close the sensor by flipping it down. Ensure that it clicks
into its closed position; see Figure 8.
A
A
Figure 8
. A: The locking mechanism on the sensor. B: The sensor in its
closed position.
7. Repeat step 1 to 6 for all reservoirs.
8. Ensure that the reservoirs are placed on a flat and
stable surface. If using the optional secondary solvent
containment, place the metal plate that can be ordered
separately (P/N 418565SP) at the bottom of the containment
(see Figure 9).
Figure 9. If using the optional secondary solvent containment, place
the metal plate that can be ordered separately at the bottom of the
containment.
6
B B
Biotage
®
Cable and Tube Routing
Link each sensor cable to its solvent/waste tube using five clips
and three markers:
1. Link the sensor cable to its solvent/waste tube using five
clips with a distance of approximately 15 cm (5.9"); see C1 to
C5 in Figure 10. Place the first clip (C1) after the heat shrink
tube; see Figure 11.
Note: Twenty (20) solvent clips, and five (5) waste clips
are included in the support kit (P/N 418956SP). See the
difference in Figure 12.
2. Link the sensor cable to both ends of the solvent/waste tube
using two markers with the correct label (S1, S2, S3, S4,
or W); see M1 to M2 in Figure 10. Place the first marker (M1)
on the heat shrink tube; see Figure 11.
3. Attach a marker with the correct label to the end of the
sensor cable; see M3 in Figure 10.
Note: Two markers per solvent/waste tube are included in
the support kit (P/N 418956SP). The third marker is supplied
with the system.
M1
M1
C1 C1
C2
C2
M3
M3
C3
C3
Figure 10
. The cable and tube routing using clips and markers with S1-S5
and W.
C1 C1
Figure 11
. Place the first marker (M1) on the heat shrink tube and the first
clip (C1) after the heat shrink tube. (The photograph is showing a close-up
of M1 and C1 in Figure 10.)
Sensor cable
Sensor cable
Solvent tube
Waste tube
Figure 12
. The solvent and waste clips.
Liquid Level Sensor for Biotage
®
M2
M2
C4
C4
M1
M1
Selekt Family – User Manual | © Biotage 2023
C5
C5