This device is an accessory to a vertical air cell cushion. It is designed to assist in the setup and maintenance of the
Cushion. It does not replace cushion manufacturer setup and maintenance processes. See the manufacturer's cushion manual
for instructions on specific cushion setup.
WARNING: DO NOT sit on the cushion when not inflated or before Sentinel is installed. It is unsafe to sit on a
vertical air-cell cushion devoid of air.
WARNING: Sentinel default levels are factory settings and should not be taken as clinical recommendations.
Sentinel is designed to maintain a clinically set level of air as set by the user's clinician and as such cannot
prevent bottoming out; a suitably experienced person should check cushion status as per the cushion's manual
to ensure the correct level of air is achieved.
WARNING: A clinician should check levels using the recommended method by the cushion manufacturer's
manual.
WARNING: Turning auditory alerts off does not cure the problem that the user is being alerted to. Ignoring an
alarm may place the user at risk. The alarm will sound again after 30 minutes if the underlying cause has not
been resolved.
WARNING: DO NOT sit on the cushion when detaching Sentinel. If user is sitting during detachment, air in the
cushion could leak rapidly out of the open fill valve and bottom out the user.
WARNING: After Sentinel is disconnected, DO NOT sit on cushion until the valve is fully closed.
• Sentinel cannot replace the need to perform skin checks and cannot prevent the cushion from bottoming out.
• Contact your clinician if you have any concerns regarding cushion pressure settings.
• Sentinel does not replace the need to perform regular pressure shifts.
WARNING: DO NOT let Sentinel dangle, it must be clipped to a chair or cushion without a kink in the hose or be
retained in a cushion pocket or via a wheelchair bracket.
WARNING: DO NOT clip Sentinel to the user.
WARNING: DO NOT use Sentinel while showering or with a shower cushion, Sentinel must be used on a cushion
with a cover properly installed according to the cushion user manual.
WARNING: DO NOT tug on Sentinel while it is in use.
WARNING: Charge with provided USB-C and wall plugs, do not use third party cables or plugs.
WARNING: DO NOT use Sentinel if there is any visible damage to the Sentinel unit or tubing.
WARNING: DO NOT use Sentinel if it has been submerged in water or exposed to heavy rain.
• Sentinel will not function in temperatures above 131 degrees F/55 degrees C.
• When not in use, Sentinel should be stored at room temperature and should not be exposed to extreme
temperatures or direct sunlight for extended periods.
WARNING: User must check Sentinel is working prior to each use.
• Sentinel needs charging. Ignoring the low battery warning will stop the device from functioning correctly.
• When fitting the hose, always ensure it is fitted fully over the valve and routed in such a way that it is not at risk
of accidental detachment or kinking.
WARNING: Sentinel contains a lithium-ion battery. These batteries can develop a thermal event if damaged.
Thermal events can cause a warming of the battery or release smoke and vapor. Never attempt to remove the
lithium ion battery from Sentinel as this can damage the battery.
WARNING: If the outlet hose is visibly damaged or if you have questions regarding the seal of the hose to the
cushion, refrain from using sentinel until the proper seal can be verified.
WARNING: Sentinel must never be cleaned by or exposed to any of the following; acetone or any other caustic
cleaners,Sunscreen, Dishwasher, Running water, Stagnant water or liquid, Bleach, Chlorine based solutions.
WARNING: Ensure the device does not apply pressure to the user's legs when placed in the pocket or clipped to
the cushion handle, as this could potentially cause pressure injuries.
WARNING: Check to make sure the device does not press into the legs when device is in the cushion pocket
WARNING: In the case of any serious incident occurring in relation to the device, this should be reported to the
manufacturer and the competent authority of the Member State in which the user and/or patient is established.
WARNING: While in clinician mode, ensure the hose is not obstructed or bent in a way to prevent air flow.
WARNING: While using clinician mode, the user must remain seated until clinician mode is exited.
WARNING: DO NOT sit on the cushion while Sentinel is connected to the cushion and charging. Always ensure
the device is disconnected from its charger during use.
WARNING: Do not modify, alter, or disassemble Sentinel outside of modifying the hose length.
WARNINGS
9