Descargar Imprimir esta página

Therabody THERAGUN RELIEF Manual página 11

Ocultar thumbs Ver también para THERAGUN RELIEF:

Publicidad

Guidance and manufacturer's declaration – electromagnetic immunity – for all EQUIPMENT and SYSTEMS
Guidance and manufacturer´s declaration – electromagnetic immunity
The Theragun RELIEF is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of the Theragun RELIEF should assure that it is used in such an environment.
Immunity test
Electrostaticdischarge (ESD)
IEC 61000-4-2
Electrostatic transient / burst
IEC 61000-4-4
Surge
IEC 61000-4-5
IEC 60601 test level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
Compliance
level
Floors should be wood,
± 8 kV contact
concrete or ceramic tile.
If floors are covered with
± 2 kV, ± 4 kV, ± 8 kV,
synthetic material, the
± 15 kV air
relative humidity should be
at least 30 %.
Mains power quality should be
N/A
that of a typical commercial or
home environment.
Mains power quality should be
N/A
that of a typical commercial or
home environment.
Electromagnetic envi-
ronment -guidance
EN
11

Publicidad

loading