Softform®Premier Active 2 Mattress
WARNING Risk of entrapment!
Patient entrapment with the bed side rails may cause injury or
death. A thorough patient assessment should be completed
and monitored and the equipment should be used as specified
and maintained to reduce the risk of entrapment. Variations in
bed rail dimensions, and mattress thickness, size and density
could increase the risk of entrapment.
▸ Mattress must fit bed frame and side rails to prevent patient
entrapment. Follow the bed manufacturer's instructions.
▸ After any adjustments, repair or service and before use,
make sure all attaching hardware is tightened securely.
Rails with dimensions different from the original equipment
supplied or specified by the bed manufacturer may not be
interchangeable and may result in entrapment or other
injury.
WARNING
The
Softform®Premier
talled on medical bed frames with bed sides or assist rails. It is
preferred that the rails to be in the raised position whenever
the patient is on the bed. Healthcare professionals assigned
to each case should make the final decision as to whether
side assist rails are warranted after assessing patient risk of
entrapment.
Controls on the footboard may be obstructed by the control
unit on a few bed frames. It may be necessary to relocate the
control unit.
▸ Before placing the patient on the bed, check that air hoses
and power cord are clear of moving bed components.
▸ Operate all bed frame motorized functions through their
full range of motion to be certain that there is no pulling,
interference or pinching.
8
Active 2 is recommended to be ins-
4.2 Installing the system
1. Place the control unit at the end of the
bed (preferably hung by the built in
control unit hooks).
2. Detach the
twin hose from the end of the mattress.
3. Connect it to the control unit.
(An automatic click will signify a secure
connection).
4. Connect the mains power lead to the
control unit and a suitable power supply.
5. Switch on mains power.
6. Switch on control unit switch.
System will pressurise (see chart 5.4,
shown by a green light and three audible
beeps.
Correct pressure shown by a green light and
two audible beeps.
4.3 Shut down procedure
1. Perform above in reverse.
Setup and installation
Softform®Premier
Active 2
page11)