Descargar Imprimir esta página

VenenWalker Basic Manual De Instrucciones página 81

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 103
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker emis-
sions
IEC 61000-3-3
Immunity Test standard
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient/
burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, interrup-
tions and variations
on power supply input
lines
IEC 61000-4-11
Power frequency
(50 / 60 Hz) magnetic
field
IEC 61000-4-8
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Note: UT is the AC mains voltage prior to application of the test level.
Class A
Comply
IEC 60601-1-2 Test Level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV air
± 2 kV, for power supply
lines
± 1 kV signal input / output
100 kHz repetition
frequency
± 0.5 kV, ± 1 kV differen-
tial lines
± 0.5 kV, ± 1 kV , ± 2 kV
common mode
0 % UT; 0.5 cycle. At
0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°.
0 % UT; 1 cycle and 70
% UT; 25 / 30 cycles;
Single phase: at 0°.
0 % UT; 250 / 300 cycle
30 A/m
50Hz / 60Hz
3 V
0.15 MHz – 80 MHz
6 V in ISM and amateur
radio bands between
0.15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
Compliance Level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV air
± 2 kV, for power supply
lines
± 1 kV signal input / output
100 kHz repetition
frequency
± 0.5 kV, ± 1 kV differen-
tial lines
± 0.5 kV, ± 1 kV , ± 2 kV
common mode
0 % UT; 0.5 cycle. At
0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°.
0 % UT; 1 cycle and 70
% UT; 25 / 30 cycles;
Single phase: at 0°.
0 % UT; 250 / 300 cycle
30 A/m
50Hz / 60Hz
3 V
0.15 MHz – 80 MHz
6 V in ISM and amateur
radio bands between
0.15 MHz and 80 MHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2.7 GHz
80 % AM at 1 kHz
NL