6.3 Cooling Unit with Mouthpiece
The Cooling Unit with Mouthpiece is intended for one person and can be
used by that person several times. For proper function, the Cooling Unit with
Mouthpiece must be free of deposits. Deposits on the inside of the Cooling Unit
are caused by steam condensate.
X Check the Cooling Unit with Mouthpiece before each use and after each
treatment.
X Refer to chapter 6.3.5 „Checking the Cooling Unit Parts and the Filling
Chamber Screen", page 77, for verification criteria on whether the Coo-
ling Unit with mouthpiece must be disposed of.
6.3.1 Disassembling the Cooling Unit with Mouthpiece
X Heat up the Cooling Unit for better disassembly, as the steam condensate
can be sticky.
6 C L E A N I N G / H Y G I E N E
X Pull the Cap Lock towards the
Mouthpiece.
X Pull the Cooling Unit Cap off the
Vaporizer.
X Pull the Mouthpiece out of the
Cooling Unit Cap while slightly tilting
it back and forth.
X Pull off the Mouthpiece Seal Ring.
72