Indicator Lights
Each number represents a preset cycle. Above each cycle number is an LED indicator light. The LED indicator light
changes to indicate the cycle current selected.
(1) Chemistry
(2) Coag (PPP)
(3) Urine
Cycle Selection
Lock Cycle Selection
Unlock Cycle Selection
Drucker Diagnostics – Customer Service: +1-814-692-7661 –
This setting is factory preset for Chemistry tubes
This setting is factory preset for Coagulation or Platelet Poor Plasma (PPP)
This setting is factory preset for urine tubes.
Begins running the cycle indicated by the cycle indicator LED light.
Start
The lid must be closed.
Allows access into the rotor chamber by engaging the unlocking
Unlock
mechanism. Entry is only possible when the rotor is stopped.
Pressing the UNLOCK button during operation will terminate the run
Stop
and unlock the lid after the rotor has come to a stop.
The LED light is on for the cycle currently selected. To change the
selected cycle, open lid and press the UNLOCK button in succession
until the desired cycle is selected. Two seconds after selection, the
button reverts to its UNLOCK function.
Open Lid. Select desired cycle. Press and hold the UNLOCK button
for 5 seconds. Two beeps will confirm that cycle selection is locked.
To re-enable cycle selection, open lid, press and hold the UNLOCK
button for 5 seconds. Three beeps will confirm that cycle selection
is now unlocked.