General; Description; Specifications - Promax GF-941 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

USER'S MANUAL GF-941

1. GENERAL

1.1 Description

The GF-941 Function Generator covers a wide frequency range, from 0.2 Hz to 20
MHz. It is a very adaptable device through the several variation means of the three main
waveforms it generates: sine, triangular and square.
Time symmetry can be changed and overlapping a variable continuous voltage as
well.
Two generators are combined into one. One of them is used as an auxiliary
generator for both AM and FM modulation functions, sweep function or for triggering the
main generator, supplying a given number of selected waveform periods, or burst function,
in a recurrent way. Both frequency and the auxiliary waveform can be selected at the
panel. The main generator is provided with a frequency indicator, to lessen reading errors.

1.2 Specifications

GENERAL
Output signals
Functions
FRECUENCY
Range
Frequency control
Display
Accuracy
OUTPUT
Amplitude
Output impedance
Amplitude control
Attenuator
SYMMETRY
04/2003
FUNCTION GENERATOR
GF-941
Sine, triangular or square.
Variable symmetry
AM modulation
FM modulation
Sweep
Burst
0.2 Hz to 20 MHz in 8 decades
Continuous in each decade 20:1
3½ digits digital display of the selected value in all
decades.
± 3% of measurement ± 1 digit (0,2 Hz to 15 MHz)
20 Vpp in open circuit, 10 Vpp (50Ω) up to 13 MHz
10 Vpp in open circuit, 5 Vpp (50Ω) up to 20 MHz
50Ω
Variable continuously, >20 dB
Up to 63 dB
Three selectable cells 3, 20 and 40 dB
20% to 80% continuously variable (up to 1 MHz)