Descargar Imprimir esta página

KPS -CC840 Manual De Usuario página 43

Localizador de cable

Publicidad

Idiomas disponibles

Idiomas disponibles

Cable locator
3. Please do not try to disassemble the battery! The electrolyte it contains
shows high basicity, which can cause corrosion! If the electrolyte comes into
contact with skin or clothing, use fresh water to clean the important parts. If
electrolyte gets into your eyes, use cool water to wash your eyes immediately,
and visit a doctor as soon as possible.
Caution!
1. Before changing the battery, the equipment must be switched off, all
connected measuring circuits must be interrupted, and all lead wires must be
removed.
2. Only the battery specified in the technical parameters can be used.
3. If the equipment is not going to be used for a long time, remove the battery.
If the detector is contaminated due to a battery leak, send the device to the
factory for cleaning and testing.
4. For the disposal of the used battery, please observe the existing regulations
for the recovery, reuse and disposal of batteries.
6.5 Calibration interval
To ensure the accuracy of the measurement made by the equipment, it must
be regularly calibrated by the company's adjustment personnel. The
recommended calibration frequency is annual. If the equipment is used
frequently or the conditions of use are very poor, the calibration interval will be
shorter. If the equipment is rarely used, the calibration frequency can be
extended to three years.
43

Publicidad

loading