MAINTENANCE AND USE MANUAL
C
oDe anD meaning
AL0210
High temp. in T2 probe
AL0211
Low temp. in T2 probe
during sterilisation
AL0300
P probe encoding error
AL0301
OPEN P
AL0302
P short circuit
AL0310
High pressure during
sterilisation
AL0311
Low pressure during
sterilisation
AL0400
TVP probe encoding error
AL0401
OPEN TVP
AL0402
TVP short circuit
AL0404
Low temp. in TVP probe
AL0405
High temp. in TVP probe
AL0500
TRS probe encoding error
AL0501
OPEN TRS
AL0502
TRS short circuit
AL0504
Low temp. in TRS probe
AL0505
High temp. in TRS probe
AL0600
Excessive drop during the
VACUUM TEST stabilisation
phase
AL0601
Excessive drop during the
VACUUM TEST
maintenance phase
AL0700
T1/T2 comparison
zetaclave B
EN
p
roblem
T2 probe has exceeded the preset cycle
temperature.
T2 probe detects excessively low values
during sterilisation.
Alarm due to PCB self-diagnosis.
System detects that P probe is open.
System detects short-circuit in P probe.
P probe detects excessively high values
during sterilisation.
P probe detects excessively low values
during sterilisation.
Alarm due to PCB self-diagnosis.
System detects that TVP probe is open.
System detects short-circuit in TVP
probe.
TVP probe does not reach operating
temperature.
TVP probe has reached a temperature
higher than operating threshold.
Alarm due to PCB self-diagnosis.
System detects that TRS probe is open.
System detects short-circuit in TRS
probe.
Temperature does not reach operating
temperature.
TRS probe has reached a temperature
higher than operating threshold.
An excessive pressure drop occurred
during the first 5 minutes of the
stabilisation phase, while performing
the VACUUM TEST cycle.
An excessive pressure drop occurred
during the first 10 minutes of the
maintenance phase, while performing
the VACUUM TEST cycle.
Conflicting temperature values detected
by the two interior sensors during
sterilisation.
s
olution
Reset alarm =
START/STOP + SELECT for 5
Reset the alarm then wait for 10 minutes with the
door
open.
Repeat
the
problem continues contact the after sales
service.
Reset the alarm, switch the machine off and on: if
the problem persists, contact technical support.
Reset the alarm, remove the tray holder, clean the
sterilisation chamber inside and repeat the cycle.
Reset the alarm, clean the gasket and repeat the
cycle: if the problem continues call the after sales
service.
Reset the alarm, switch the machine off and on: if
the problem persists, contact technical support.
Reset the alarm and repeat the cycle. If the pro-
blem persists, contact technical support.
Reset the alarm then wait for 10 minutes with the
door open. Repeat the cycle: if the problem conti-
nues
contact
the
after
Reset the alarm, switch the machine off and on: if
the problem persists, contact technical support.
Reset the alarm and repeat the cycle. If the pro-
blem persists, contact technical support.
Reset the alarm then wait for 10 minutes with
the door open. Repeat the cycle: if the problem
continues contact the after sales service.
Take the tray holder out: clean and
thoroughly the sterilisation chamber. Repeat the
cycle: if the problem continues you have to do a
Vacuum test cycle when the autoclave is cold.
Reset the alarm, fill the thank until the
maximum icons appears on screen and repeat
the cycle: if the problem persists in the
following cycles, contact technical support.
.
seC
cycle:
if
the
sales
service.
dry
171 of 228