Drive DeVilbiss iGo 2 125 Guía De Instrucciones página 5

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

WARNING
• Position your unit near an electrical outlet at least 6 inches (16 cm) from walls, draperies, or any other objects that might prevent the proper flow of air in and out of your
device. The iGo2 concentrator should be located so as to avoid pollutants or fumes, and placed in a well-ventilated place so that the air inlet and exhaust are not
blocked. Do not cover unit with a blanket, towel, quilt, or other covering, as the unit may overheat.
• Do not lubricate fittings, connections, tubing, or other accessories of the oxygen concentrator to avoid the risk of fire and burns. No lubricants are recommended for use
on this device.
• Use only water-based lotions or salves that are oxygen-compatible before and during oxygen therapy. Never use petroleum or oil-based lotions or salves to avoid the
risk of fire and burns.
• The Drive DeVilbiss iGo2 portable oxygen concentrators are equipped with a fire mitigating outlet fitting that prevents propagation of fire into the unit.
• In order to prevent a fire propagating from the patient through the cannula towards the unit, a means of protection should be located as close to the patient as
practicable. Country Standards may vary. Please contact your provider for information.
• Improper use of the power cord and plugs can cause a burn, fire, or other electric shock hazards. Do not use the unit if the power cord is damaged.
• Locate oxygen tubing and power supply cords to prevent tripping hazards and reduce the possibility of entanglement or strangulation.
• Use only spare parts recommended by the manufacturer to ensure proper function and to avoid the risk of fire and burns.
• Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.
• When device is used under extreme operating conditions, the temperature near the exhaust vents on the bottom of the unit may reach 57˚C (134.6˚F). Keep bare skin
away from this area.
• The following surface temperatures may exceed 41˚C (105.8˚F) under extreme conditions:
• External surface of POC .................................... 52˚C (125.6˚F)
• Exhaust gas at discharge port ........................... 53˚C (127.4˚F)
Keep bare skin away from this area.
WARNING
• If you feel discomfort or are experiencing a medical emergency while undergoing oxygen therapy, seek medical assistance immediately to avoid harm.
• Geriatric, pediatric or any other patient unable to communicate discomfort can require additional monitoring and/or a distributed alarm system to convey the information
about the discomfort and/or the medical urgency to the responsible caregiver to avoid harm.
• Use of this device at an altitude above 3000 meters (9843 feet) or above a temperature of 35˚C (95˚F) or greater than 93% relative humidity is expected to adversely
affect the flow rate and the percentage of oxygen and consequently the quality of the therapy. Refer to specifications for details regarding parameters tested.
• The oxygen delivery setting has to be determined for each patient individually with the configuration of the equipment to be used, including accessories. It is very
important to follow the prescription determined by your physician.
• Your delivery settings of the oxygen concentrator should be periodically reassessed for the effectiveness of therapy.
• To ensure you receive the therapeutic amount of oxygen delivery according to your medical condition, the iGo2 oxygen concentrator must:
be used only after one or more settings have been individually determined or prescribed for you at your specific activity levels.
be used with the specific combination of parts and accessories that are in line with the specification of the concentrator manufacturer and that were used while your
settings were determined.
• The settings of this iGo2 portable oxygen concentrator do not correspond with continuous flow oxygen device settings.
• The setting of other models or brands of oxygen therapy equipment do not correspond with the settings of this iGo2 portable oxygen concentrator.
WARNING
The proper placement and positioning of the prongs of the nasal cannula in the nose is critical to the amount of oxygen delivered to the respiratory system of the patient.
Some respiratory efforts of the patient might not trigger the conserving equipment.
Wind or strong draughts can adversely affect accurate delivery of oxygen therapy.
This device is not intended for use with a tracheotomised patient.
WARNING
To avoid electric shock, do not remove the concentrator cover. The cover should only be removed by a qualified Drive DeVilbiss technician. Do not apply liquid directly to
the cover or utilize any petroleum-based solvents or cleaning agents.
Before attempting any cleaning procedures, turn the unit "Off" and disconnect from AC or DC power.
Do not service or clean this device while in use with a Patient.
Use no lubricants, oils or grease.
Use of harsh chemicals (including alcohol) is not recommended. If bactericidal cleaning is required, a non-alcohol based product should be used to avoid inadvertent
damage.
WARNING
• This device contains electrical and/or electronic equipment. Follow local governing ordinances and recycling plans regarding disposal of device components.
MR Unsafe
• Do not bring the device or accessories into a Magnetic Resonance (MR) environment as it may cause unacceptable risk to the patient or damage to the iGo2 or MR
medical devices. The device and accessories have not been evaluated for safety in an MR environment.
• Do not use the device or accessories in an environment with electromagnetic equipment such as CT scanners, Diathermy, RFID and electromagnetic security systems
(metal detectors) as it may cause unacceptable risk to the patient or damage to the iGo2. Some electromagnetic sources may not be apparent, if you notice any
unexplained changes in the performance of this device, if it is making unusual or harsh sounds, disconnect the power cord and discontinue use. Contact your home
care provider.
• This device is suitable for use in home and healthcare environments except for near active HF SURGICAL EQUIPMENT and the RF shielded room of an ME SYSTEM
for magnetic resonance imaging, where the intensity of Electromagnetic DISTURBANCES is high.
• Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed to verify that they are operating normally.
• Use of accessories and cables other than those specified or provided by the manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
• Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any
part of the iGo2, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
SE-125-1
• External power supply ................................... 43˚C (109.4˚F)
• Battery Pack ................................................. 49˚C (120.2˚F)
EN - 5

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido