VBM Medizintechnik Manujet III Instrucciones De Uso página 10

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 12
gas probe
country
Germany
USA
France
All other countries are supplied without gas probe.
The professional choice and attachment of the pressure hose / gas probe rests with the user.
TEchNIcAL DATA
Oxygen source pressure:
Adjustable ventilation pressure:
Display accuracy:
Colour coded scale:
Operating pressure must not exceed 4 bar (58 psi).
Dimensions (case):
Weight with case and accessories:
ENVIRoNmENTAL coNDITIoNS
manujet III
Transport/Storage:
Operation condition:
connecting tube
See connecting tube label.
wIpE DISINFEcTIoN mANUjET III
For disinfecting the device must be wiped externally with a non-flammable disinfectant solution (e. g. Sani-Cloth
from Ecolab Deutschland GmbH). When choosing a product for disinfection, a disinfectant with appropriate ranges of
action must be used: bactericidal, yeasticidal, tuberculocidal and virucidal. Examine the products for visible contamina-
tion. If necessary repeat the disinfection. After wipe disinfection, the products have to be checked for damages (cracks,
ruptures etc.).
cAUTIoN
Do not immersed the product in liquids.
The product cannot be automatic or manual cleaned respectively sterilised.
Do not clean the glass of the gauge with aggressive agents.
If wipe disinfection is not successful, or if the product is strongly contaminated, the product must be disposed of as
specified in chapter "Disposal".
The connecting tube is for single use and must not be reprocessed. Reprocessing influences the function of the con-
necting tube. Any reuse involves the potential risk for cross contamination.
A list of compatible cleaning agents is available at www.vbm-medical.com in the download section.
mETRoLogIcAL INSpEcTIoN
The accuracy of the Manujet III Gauge must be checked every 24 months.
If there is an indication that the Manujet III Gauge cannot keep the accuracy of measurement within the 24 months, or
the metrological characteristics of the Manujet III Gauge have been affected, a metrological control has to be carried out
immediately. This has to be carried out only by a person who meet the requirements for metrological controls. The results
of the control, as well as the measured values and measurement procedures has to be documented.
After successful control, the Manujet III Gauge has to be marked with a sign which indicates (clearly and traceable) the
year of the next inspection and the person who carried out the metrological control.
gas probe
Oxygen DIN 13260
DISS O
2
NF S 90-116 O
2
- yellow
- blue
Height:
Width:
Depth:
- Manujet III Standard:
- Manujet III with NIST-O
+5 °C ... +40 °C
(41 °F - 104 °F)
+10 °C ... +40 °C (50 °F - 104 °F)
30 ... 95 % rel. atmospheric humidity without condensation
-10-
4 - 10 bar
0 - 3.5 bar
±0.05 bar
1.0 - 2.5 bar
2.5 - 4.0 bar
100 mm
400 mm
300 mm
1.80 kg
:
1.24 kg
2
(58 - 145 psi)
(0 - 51 psi)
(±0.72 psi)
(14.5 - 36 psi)
(36 - 58 psi)
(4 inch)
(16 inch)
(12 inch)
(60 oz)
(40 oz)
Active
®