Description of where the unit is to be installed (important when using more than
one unit). 20 characters.
Hint: This is also stored on the diskette. Graphics and tables are identified
Unit identifier
using this text when operating with a PC (important if more than one unit is to
be used). The unit identifier is also available when exporting data into other
spread sheet programmes.
Format dependent on operating language, see help text in the unit
Actual date
(e.g. DD.MM.YY)
Actual time
Format: hh:mm, 24 hour clock display
Activate summer/normal time changeover.
Summer time
"Automatic": Change to the valid regulations for the selected NT/ST region
changeover
"Manual": Changeover times can be set up in the following addresses
"Off": No changeover times needed
The changeover times for normal to summer time vary dependent on
NT/ST region
geographic region. Select the required region, for control purposes the
changeover times are displayed in grey in the next addresses.
Date NT-ST
Day, when changeover from normal time to summer time is to occur.
Time at which the internal clock is to increase by 1 hour on changeover from
Time NT-ST
normal to summer time.
Format: hh:mm
Date ST-NT
Day, when a return from summer time to normal time is to occur.
Time at which the internal clock is to be turned back by 1 hour on changeover
Time ST-NT
from summer to normal time.
Format: hh:mm
Factory default: "0000", this means changes are possible without the need of a
security code. Individual code set: Further setting up of the unit can only be
Release code
done after entering this code.
Hint: Note the code and store out of reach of unauthorised personnel.
Selection of the required temperature units. All directly connected
Temperature units
thermocouples or RTDs are displayed and plotted in these preset units.
Indicates how fast the recording is updated or stored. Is always compared to
Standard feed rate
the paper feed of a normal recorder (typically 20 mm/h).
Feed rate in alarm condition, when analogue measurements drift outside preset
Alarm feed rate
alarm set points. Selectable for each analogue input under "Alarm set point -
Feed rate".
Measurement point identifiers can contain a max 10 characters. Selecting "With
additional channel info" enables entry of 13 character additional channel
infomation per channel (e.g. respective measurement point number, tag
Channel identifier
number systems in power stations etc.).
Hint: In normal operation the additional info can be listed together with the 10
character channel identifier using the = push buttons (see 5.
"Functions/handling in operation").
• Picture: Settings for measured value plot display.
4. Setting up unit - SET UP
47