Descargar Imprimir esta página

Medivon Smart EMS Manual Del Usuario página 34

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 58
RF emission
CISPR 11
Harmonic
emissions IEC
61000-3-2
Voltage
fluctuations/
flicker
emissionsIE
C 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The PG-2601B8 is compliance for each EMISSIONS test specified
by the standard, e.g. EMISSIONS class and group
Immunity test
Electrostatic
discharge (ESD) IEC
61000-4-2
Radiated RF EM
fieldsIEC
61000-4-3
Electrical fast
transient/burst
IEC 61000-4-4
SurgeIEC
61000-4-5
Conducted
disturbance
sinduced by
RF fields
IEC 61000-4-6
Class B
Class A
Complies
IEC 60601-1-2
test level
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
10V/m
80MHz-2.7GHz
80% AM at 1kHz
±2 kV
100 kHz repetition
frequency
±0.5 kV, ±1 kV line-to-line;
±0.5 kV, ±1 kV and ±2 kV
line-to-ground;
3V 0.15MHz-80MHz
6V in ISM and amateur
radio bands between
0.15MHz and
80MHz 80% AM at 1kHz
The PG-2601B8 is
suitable for use in all
establishments,
including domestic
establishments and
those directly connected
to the public low-voltage
power supply network
that supplies buildings
used for domestic
purposes.
Compliance level
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
10V/m
80MHz-2.7GHz
80% AM at 1kHz
±2 kV
100 kHz repetition
frequency
±0.5 kV, ±1 kV line-to-line;
±0.5 kV, ±1 kV, and ±2 kV
line-to-ground;
3V 0.15MHz-80MHz
6V in ISM and amateur
radio bands between
0.15MHz and 80MHz
80% AM at 1kHz

Publicidad

loading

Productos relacionados para Medivon Smart EMS

Este manual también es adecuado para:

Mvo00925