Main Menu; Starting A Titration - Xylem SI Analytics TitroLine 7500 KF trace Manual De Instrucciones

Ocultar thumbs Ver también para SI Analytics TitroLine 7500 KF trace:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

104
3.5

Main Menu

After power-up, the main menu is always the first menu to appear.
The method displayed will always be the last method that was used (Fig. 23).
Fig. 23

Starting a Titration

3.5.1
Once all preparations have been finished, you can start to titrate samples. The titrator starts automatically with
the conditioning process when it switched on and anolyte is present in the titration vessel. Here the titration cell is
titrated until it is dry. This includes the entire water of the anolyte and even adherent liquids in the titration vessel.
If the start drift value is not reached (normally 10 µg/min) the method can´t be started.
Start the selected method with <START> in the main menu. If the drift value is below a set value (normally 10
µg/min) you will be prompted to dose the sample (Fig. 24), for the sample identification (Fig. 25) and the
weighed-in quantity (Fig. 26).
Fig. 24
Fig. 25

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

285220860

Tabla de contenido