Descargar Imprimir esta página

Xylem SI Analytics TitroLine 7500 KF trace Manual De Instrucciones página 87

Ocultar thumbs Ver también para SI Analytics TitroLine 7500 KF trace:

Publicidad

Idiomas disponibles

Idiomas disponibles

Ethernet Interface:
Stirrer/pump:
Housing:
Material:
Front keyboard:
Dimensions:
Weight:
Ambient conditions:
Climate:
Altitude:
Pollution degree:
Software
Measuring range:
Measuring speed:
Number methods:
Conditioning:
End point criteria:
Auto start:
Statistics:
Curve printout:
Documentation:
for connecting a local network (LAN)
12V DC out, 500 mA
power supply for stirrer TM 235 and KF titration stand TM 235 KF
Polypropylene
polyester coated
15.3 x 45 x 29.6 cm (W x H x D)
approx. 2.3 kg for basic unit without stirrer TM 235 or KF titration stand TM 235 KF
Do not use the device in hazardous locations!
Ambient temperature: + 10 ... + 40 °C for operation and storage
Humidity according to EN 61 010, Part 1:
Max. relative humidity 80 % for temperatures up to 31 °C,
linear decrease down to 50 % relative humidity at a temperature of 40 °C
Device: No restrictions
Power supply: up to 5000 m
Pollution degree IP 20, indoor use only.
10 µg – 100 mg / 1 ppm – 5 % (recommended)
Max. 1.5 mg/min
50
Automatically with drift determination
Drift, stop drift tolerance, min. and maximum titration time
Yes, after sample allowance
Mean value, standard deviation and relative standard deviation
Measuring unit/time
GLP compliant on printer or in PDF format (USB- memory stick)
87

Publicidad

loading

Este manual también es adecuado para:

285220860