Gima THP59J Manual De Uso Y Mantenimiento página 10

Termómetro infrarrojo de oído profesional
Ocultar thumbs Ver también para THP59J:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Specifications
- Temperature measurement range: 34~42.2°C (93.2~108°F)
- Operating temperature range: 10~40°C (50~104°F), 15%~85% RH
- Storage temperature range: It should be stored at room temperature between -20~+50°C
, RH ≤ 85%
- Transportation temperature shall be less than 70°C, RH ≤ 95%
- Comply with ASTM E1965-98, EN ISO 80601-2-56, IEC/EN60601-1-2(EMC), IEC/EN60601-
1(Safety) standards, ISO10993, RoHS.
- Accuracy: +/-0.2°C (0.4°F) within 35~42°C (95~107.6°F), +/-0.3°C (0.5°F) for other range.
- Fever alarm & Memory, °C / °F Switch function
- Battery: one lithium cell battery (CR2032 *1pcs). Battery life is estimated at 3,000 continuos
readings. Enable normal use: 1 year
- This thermometer converts the ear temperature to display its "oral equivalent." (according
to the result of the clinical evaluation to get the offset value).
- There is no gender and age limitation for using infrared thermometer.
- This is not an AP or APG product.
Normal body temperature
Normal body temperature is a range that will fluctuate throughout the day. The normal range
for adult body temperature is typically considered to be 36.1 to 37.8° C (97 to 100°F). The
body temperature varies by age, person, gender, time of day, position of body, and is usually
highest in the evening. It can be affected by activity, emotion, clothing, medications, ambient
conditions, hormones and other factors. Normal temperature will also be different depending
on the place on the body at which the temperature is taken, with rectal or ear temperature
readings generally being higher than oral temperature readings, and armpit temperature
readings generally lower than oral readings.
Temperature variations affected by age:
Young
0 ~ 2 years:
3 ~ 10 years: 36.1°C (97.0°F) ~ 37.8°C (100.0°F)
11 ~ 65 years: 35.9°C (96.6°F) ~ 37.6°C (99.7°F)
> 65 years:
Clinical repeatability: 0.18°C (<1 year old), 0.12°C (1~5 years old), 0.10°C
Old
(>5 years old)
The "Clinical Bias" is -0.2 ~ -0.4°C.
The "Limits of Agreement" is 0.49.
The "Repeatability" is 0.14°C.
ENGLISH
36.4°C (97.5°F) ~ 38.0°C (100.4°F)
35.8°C (96.4°F) ~ 37.5°C (99.5°F)
10
High
Low

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

25575

Tabla de contenido