MD 9270 Leakage Clamp TRMS Meter with Power Function Technical Specifications
Active Power
Function
W, VAr, VA
KW, KVAr, KVA
PF
phase
Note: For sinwave input minimal measured value is 1 % of measuring range.
Range
0.000 ... 9.999
10.00 ... 99.99
100.0 ... 999.9
1000 ... 9999
0.000 ... 9.999
10.00 ... 99.99
100.0 ... 999.9
0.00 ... 1.00
-180.0 ... 180.0
Resolution
0.001
0.01
0.1
1
0.001
0.01
0.1
0.01
0.1
34
Accuracy
± 1% of reading ± 30 digits
± 1% of reading ± 30 digits
± 1% of reading ± 30 digits
± 1% of reading ± 30 digits
± 2% of reading ± 30 digits
± 2% of reading ± 30 digits
± 2% of reading ± 30 digits
± 1% of reading ± 1 digit
± 1% of reading ± 4 digits