- If the carriage is positioned
within the programmed range
of motion at the time therapy
begins, the therapy session will
start immediately.
If the carriage is positioned
outside the programmed range
of motion at the time therapy
starts, it will first enter the position
"extension +10°". The carriage
will stop in this position and you
can initiate the therapy session
by pressing the START/STOP
key again.
5.2 Information about
the ARTROMOT®-K1
classic therapy
parameters
- You select a function by pressing the
corresponding control
- You change the treatment values by
turning the depressed control.
- You initiate the treatment session by
pressing the START/STOP key
I Extension (streching)
- Maximum knee extension:
-10 degrees
- Maximum hip extension:
7 degrees
I Flexion (bending)
- Maximum knee flexion:
120 degrees
- Maximum hip flexion:
115 degrees
Note!
The programmed value and the
value measured at the patient's
knee may deviate slightly.
I Speed
The speed can be adjusted between
5 % and 100 % in steps of 5 %.
Default setting: 50 %
I Pauses
Pauses occur at the selected limits
where stretching turns into bending
and bending into stretching (selected
extension and flexion values).
The value entered applies to both
extension and flexion pause.
Pauses can be set to any value bet-
ween 0 and 59 seconds in steps of 1
second, and to values between 1
and 59 minutes in steps of 1 minute.
When selecting the seconds, the
value changes in 1-second steps for
the first five seconds. Subsequently
the interval changes to 5-second
steps.
The minutes are always adjusted in
1-minute steps.
Default setting: no pause
41