as they may strangle themselves with the patient's tube and/or the power cable.
12.
The medical device is in contact with the patient by means of a disposable probe (supplied with the device) furnished with
the relative CE compliance certification according to the requirements of regulation ISO 10993-1: thus, no allergic reactions
and skin irritations may occur.
13.
The product and its parts are biocompatible in accordance with the requirements of regulation EN 60601-1.
14.
Operation of the device is very simple and therefore no further explanations are required other than those indicated in the
following user manual.
The manufactured cannot be held liable for accidental or indirect damages should the device be modified,
repaired without authorization or should any of its component be damaged due to accident or misuse.
Any minimal modification / repair on the device voids the warranty and does not guarantee the compliance
with the technical requirements provided by the MDD 93/42/EEC (and subsequent changes) and its
Before using the AS-200, consult the instructions for use: failure to read all the instructions in this manual can be harmful
-
for the patient.
Don't use the device for thoracic drainage.
-
The device must not be used for suction of explosive, corrosive or easily flammable liquids.
-
AS-200 is not suitable for MRI. Do not introduce the device in MRI environments.
-
Model
Typology (MDD 93/42/EEC)
Classification UNI EN ISO 10079-1
Main Voltage
Power consuption
Fuse
Maximum suction aspiration (without jar)
Maximum flow (without jar)
Weight
Dimension
Functioning
Working condition
Conservation and transport condition
Use a soft dry cloth with not – abrasive and not – solvent detergents. To clean the device external parts always use a cotton cloth
dampened with detergent. Don't use abrasive or solvent detergents.
PARTICULAR CARE SHOULD BE TAKEN TO ENSURE THAT THE INTERNAL PARTS OF THE EQUIPMENT DO
NOT GET IN TOUCH WITH LIQUIDS. NEVER CLEAN THE EQUIPMENT WITH WATER.
During all clearing operations use protection gloves and apron (if need be, also wear a face mask and glasses) to avoid getting in
contact with contaminating substances (after each utilization cycle of the machine).
normatives.
CONTRAINDICATIONS
TECHNICAL CHARACTERISTICS
Regolable from -75kPa (-0.75 Bar) to -10kPa (-0.10 Bar)
Room temperature:
Room humidity percentage:
Atmospheric pressure:
Room temperature:
Room humidity percentage:
Atmospheric pressure:
CLEANING DEVICE
13
AS-200 (REF RE 410120)
Medical Device Class IIa
HIGH VACUUM / LOW FLOW
230 V ~ / 50 Hz
184 VA
F 1 x 1.6A L 250 V
-75kPa (- 0.75 bar)
16 l/min
2.5 Kg
357 x 193 x 215mm
NON-STOP OPERATED
5 ÷ 35 °C
10 ÷93 % RH
700 ÷ 1060 hPa
- 25 ÷70 °C
0 ÷ 93% RH
500 ÷ 1060 hPa