The device must be checked before each use in order to detect malfunctions and / or damage caused by transport and / or
storage.
The working position must be such as to allow one to reach the control panel and to have a good view of the empty
indicator, the jar and the antibacterial filter.
It is recommended not to keep the device in your hands and / or to avoid prolonged contact with the body of apparatus.
WARNING: For proper use, place the aspirator on a flat, stable surface in order to have the full volume of use of the jar and
better efficiency of the overflow device.
Place the unit on a flat, horizontal surface
Connect one end of the short silicon tube, with antibacterial filter, to the suction connector on the lid of the jar.
The other tube already connected to the filter has to be connected to the "VACUUM" jar outlet, where has been fixed the red
float (security float). When the 90% of the volume of the jar is reached there is the activation of the security float (the float
close the aspiration connector on the jar) to avoid liquid penetration inside the device.
WARNING: Ensure that the FLUID SIDE or IN marker on the filter is on the side facing the collection jar lid and fitted into the
"VACUUM". A wrong connection causes immediate destruction in case of contact with sucked liquids. (SEE PAG. 16)
Filter assembling
Suction Pump
Inlet
Connect the long silicone tube to the "PATIENT" jar outlet
Connect the other end of the long silicon tube to the probe plastic connector, then connect the suction probe to it.
Connect the power cord to the device then connect the plug to the electrical mains supply.
Push switch on position I to start suction.
Unscrew the lid of the jar and fill the jar 1/3 full or ordinary water (this assists the unit to reach peak vacuum performance
and makes clean-up easier) then re-screw the lid on the jar correctly.
During operation the jar has to be in vertical position to avoid overflow valve to cut off aspiration. Should this happen,
switch off the device and disconnect the tube from the jar cover (from "VACUUM" outlet).
Once finished push switch on O position and unplug.
Remove the accessories and clean.
At the end of each use, place the device in its box away from dust.
WARNING: The power supply cable plug is the element of separation from the electrical mains system: even if the units
equipped with a special on / off switch button, the power supply plug must be kept accessible once the device is in use so as to
allow a further method of disconnection from the mains supply system.
INSTRUCTION FOR USE
FLOW DIRECTION
IN
17
IN / Fluid Inside
Jar Air Tube