TROUBLESHOOTING (Continued)
If an error message is displayed, refer to the following list for possible corrective actions you can take.
If these do not correct the problem, contact your dealer.

ERROR MESSAGE

"Err 2"
"Err 3"
"Err 5"
"Err 12"
"Err 7"
"Err 10"
"Err 11"
"Err 13"
"Err 14"
All other errors

CLEANING

Keep the drive enclosure clean with mild detergents. Do not immerse.
REPLACEMENT PARTS & ACCESSORIES
77500-24
07595-43
77300-32
®
CAUSE
Motor over-speed
Overload
Bad data Operator
parameters set to
default values
Voltage out of range
Over temperature
Internal error or failure
Fuse—T6.3A 5 x 20 mm
Footswitch
Remote control cable, 25 ft (7.62 m)
REMEDY
1 Clear by pressing any key
2 Check for proper tube loading and pump operation
3 Return unit for repair if the error persists
1 Clear by pressing any key
2 Check for proper tube loading and pump operation
3 Return unit for repair if the error persists
1 Clear by pressing any key
2 Reprogram operator parameters
3 Return unit for repair if the error persists
1 Clear by pressing any key
2 Check that AC line voltage is within specified
voltage ranges
3 Return unit for repair if AC line voltage is correct
and the error persists
1 Check for heat sources or obstructions to cooling
2 Check for proper tube loading and pump operation
3 Allow unit to cool Clear by pressing any key
4 Return unit for repair if no cause for overheating
is found and the error persists
1 Clear (if possible) by turning power off and on
2 Return unit for repair if the error persists
8
®