3. Operations in uploading
1) System setting
Function:
• If there's much noise in the wave, you can set "Filter Settings" in "Sample
Setting" to get calm and clear ECG.
• If there's much noise when sampling, you are supposed to select "PF Filter"
in "Sample Setting".
• Keep off interference, refer to "Filter Settings" to solve noise and baseline
drift.
• If myoelctrical noise is serious you can select "myoelctrical Filter ".
• You can use the "Alarm Settings" to set the scope of heart rate, when alarm
will become red.
2) Stop
Function: Stop the ongoing data uploading.
Gain and speed
Function: You can adjust the waveform gain and walking speed.
Amplitude: 5mm/mv, 10mm/mv, 20mm/mv
Velocity: 12.5mm/s, 25mm/s, 50mm/s
Note: After uploading, the computer and devices can be disconnected.
6.4 Review
You can review the local case info and print report by this function.
The main interface of "Review" is shown as Fig. 6-16 (ECG Case) or Fig. 6-17
(SpO
Case).
2
Fig. 6-16
ENGLISH
Query condition
Case Table
Modify case
74