Fig. 42
Stage 5 corresponds to maximum titration speed. Speed is reduced by 50% each time.
Example: 20 dosing unit:
Stage 5
100 % (ca. 40 ml/min)
Stage 4
50% (ca. 20 ml/min)
Stage 3
25% (ca. 10 ml/min)
Stage 2
12.5 % (ca. 5 ml/min)
Stage 1
6.8 % (ca. 2.5 ml/min)
As soon as the titration is completed, press the <STOP/F2> key or approx. for 1 sec. the grey key of
the "mouse". The titration result will be calculated and displayed.
Fig. 43
The result can also be printed or stored in PDF- and CSV-format. <ESC> will take you back to the
start menu way to start the next titration immediately. Filling of the dosing unit occurs automatically.