The table below outlines these additional troubleshooting issues:
TC is high and FAC
is zero or low
TC is reading higher
than FAC
FAC is reading high
6.4

Flow Meter

Caution: Prior to opening the controller or installing any electrical components, turn off all power
supplies to the controller

6.4.1 Selecting and Location

The exact location of the flow meter will depend on the specific piping arrangement. Several factors should
be considered in order to provide maximum safety and accuracy.
Safety: In order to provide prevent chemicals from being added when there is no flow to the pool, the flow
meter should be installed in the main return line after the filter.
Accuracy: in order to ensure accuracy, the flow meter should be installed in a straight run of pipe following
the minimum requirements listed in Table 10.
Table 10: Minimum Installation Distance from Pipe Fittings
Chlorine Level is too high
(~20ppm) and is bleaching
reagents
FAC was calibrated but TC
did not increase or
decrease as expected
DPD3 is being trapped on
piston
Fitting
Flange
90° elbow
Reducer
Valve
Outlet
Dilute and test manually -- controller
will operate fine once back in normal
level
a) Perform system reset and wait for
next TC reading
b) upgrade to new software version
V4.0 or later
Replace silicon piston with new
version
Distance to Flow meter
10 x ID
25 x ID
15 x ID
50 x ID
5 x ID
63