9.5
Tables / Charts
CYCLES
121° C
(wrapped and
unwrapped)
134° C
(wrapped and
unwrapped)
134° C
Flash
134° C
Prion
134° C
Helix / Bowie&Dick
Vacuum
The pre-heating time and fractionation of the vacuum can vary depending on the
conditions of the autoclave from 25 - 35 minutes that will add up with the times of the
cycles shown in table. All sterilization cycles have 3 stages of vacuum.
Dynamic sterilizer chambre pressure
Air Leakage
Empty chamber
Solid load
Small porous items
Small porous loads
Full porous load
Hollow load B
Hollow load A
Multiple wrapping
Dryness, solid load
Dryness, porous load
* in compliance with all applicable clauses of standard
STERILIZATION TABLE
DRYING
EXPOSED
TIME T4
T5 (Minutes)
(Minutes)
18L
18
12
5
12
4
3
20
12
3.5
TABLE OF TESTS TYPE AS RULE EN13060:2009
Type test
PRESS.WORK
TIME
MIN AND MAX
(relative bar)
24L
13
1.04 ÷ 1.30
13
2.02 ÷ 2.38
4
2.02 ÷ 2.38
13
2.02 ÷ 2.38
4
2.02 ÷ 2.38
-0.9
ENGLISH
TEMPERATURE
WORK MIN AND
MAX ( ° C )
121 ÷ 125
134 ÷ 138
134 ÷ 138
134 ÷ 138
134 ÷ 138
Operative cycle
X
X
X
X
X
X
X
X
X
X
X
X
29