5
Installing the apparatus
This medical device is intended for use by a qualified surgeon during endoscopic exploration.
No special training is required to install it. Kindly refer to the instructions in this manual.
Place the control unit on its base and have to hand the various accessories required to operate it.
5.1 Connections
•
Connect the power cord to the mains socket of the camera
output ports: we recommend selecting the DVI output
•
Connect the second end of the video lead to the video input port on the monitor.
•
If you wish to control certain functionalities of the insufflator or the light source, connect these to
the SOPROCOM socket on the camera's control unit
cables provided.
Connecting the endoscope to the sensor:
•
Turn the ring on the endoscope-holder on the camera head clockwise.
•
Once the endoscope is inserted, release the ring.
•
If the sensor has to be locked into position on the endoscope, the ring should be turned in the opposite
direction.
•
Connect the potential equalization cable to the equipotential plug
5.2 Activation
To optimize the quality of the images obtained with this product, always be sure to connect the
camera head before turning on the control unit.
•
Connect the power cable to the plug
•
Connect the DVI-D cable to the corresponding output port
the corresponding input port on the monitor.
èThe DVI-D output produces a progressive 1080 signal.
•
If needed, connect other monitors or peripherals, preferably using the HDSDI & 3GSDI
C8
output ports
.
èHDSDI, 3GSDI output ports produce an interleaved 1080 signal.
èHDSDI, 3GSDI output ports are preferred when a very long cable is necessary.
C2
C6
for connecting the main surgical monitor.
C9
intended for this purpose, using the two
C2
at the back of the camera.
C6
and a video lead to one of the video
C13
at the back of the apparatus.
at the back of the product and to
13