Heidolph LABOROTA 20 Instrucciones De Empleo página 88

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

E
Depress Select and Start Stop keys near Display I together and hold for 5 seconds, turn master
switch ON, release keys
„CAL" is shown in the Display Speed, „VAC" in the Display Time
750 is shown in Display I and 1 is shown in Display II (1st step).
Display I
750
1
CAL
VAC
Display Speed
Now, set the pressure measured with the separate vacuum
gauge in Display I, by pressing the Set increase key or Set
decrease key
Confirm with Start Stop key located next to Display I
Display I to show 50 and display II 2 (2
At this point, the LABOROTA 20 control automatically
opens the vacuum valve and evacuates until reaching
vacuum limit (the value on the separate vacuum gauge
won't change anymore)
In Display I, set the pressure measured with the separate
vacuum gauge in Display I, by pressing the Set increase
key or Set decrease key
Confirm with Start Stop key located next to Display I. As
confirmation for the accepted calibration values, „0"
appears in Display I and Display II
Turn master switch (Power) OFF and ON
Calibration is complete
Display II
Select key
Display Time
Taste
Start /Stop
nd
step).
88
increase
decrease

Publicidad

Tabla de contenido
loading

Tabla de contenido