Sartorius Entris Serie Instrucciones De Funcionamiento página 11

Balanzas de laboratorio
Ocultar thumbs Ver también para Entris Serie:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 65
L
R
R
L
L
R
1
Levelling the Balance
Purpose:
– To compensate for unevenness at the place of
installation
Always level the balance again any time after it has
been moved to a different location. Only the 2 front
feet are used for leveling.
Screw in both rear support feet (only on models with
t
a rectangular weighing pan).
Turn the front leveling feet as shown in the
t
illustration until the air bubble is centered within
the circle of the level indicator.
> Normally, several leveling steps are required.
On models with a rectangular weighing pan:
t
Screw out both back leveling feet until they touch
the setup surface.
Below-Balance Weighing
A port for a below-balance weighing hook is located
on the bottom of the balance.
y Not permitted for applications in legal metrology.
Lift cover plate out of the bottom of the balance.
t
Attention: Place the balance on its side, do not turn
over completely!
Secure hook 1: Use a wire, for example, to suspend
t
the sample on the hook.
y Install a draft shield if necessary.
11

Hide quick links:

Publicidad

Tabla de contenido
loading

Tabla de contenido