rotation (go by arrow marks). Sense of rotation may be changed on standstill only.
-
CAUTION: hit STOP key (32), select sense of rotation, START item
Sense of rotation is confirmed by an LED (40).
-
Start pump with „Start / Stop" key (32). Pump running is confirmed by an LED (44).
-
Hit „Start Stop" key (32) once more to stop pump.
-
When hitting the „max." key (33), pump will run at full speed as long as key (33) is held
-
depressed. This feature speeds up filling and draining of hoses without doing any changes to
the basic setting. "Full" as plain text appears in the display (41), when "max." key is
depressed.
Actual speed is displayed (41) in rpm (1/min), as long as the "speed rpm" LED is lighted.
-
To alter speed setting (+ & -), use arrow keys (35).
-
Feed rate is proportional with speed. Speed is displayed in rpm (1/min). For speed setting
-
make reference to the feed rate graphs (Specifications).
9.2. Charging / Discharging of hose
As long as key "max" (33) is pressed pump runs with maximum speed. This function is useful to
speed up the charging and dischargiung of the hose without changing the settings. When key
"max" ist pressed display shows "Full".
9.3. Extra functions
The PD 5201 and PD 5206 pump drives feature the following options:
display of feed rate in ml/minute
-
metering by volume
-
interval metering
-
calibrate feed rate or volume
-
To allow using the above options, you need to know about the pump model used; in case of
peristaltic pumps, you need to know size of hose. In as shipped condition, the SP quick peristaltic
(PU1) is preset. When using a different pump head, you need to program the item accordingly.