7.6.1. Electromagnetic compatibility:
NF EN 61326-1 (Ed. 97) + A1, industrial environment category
7.6.2. Mechanical protection
IP 53 per NF EN 60529 (Ed. 92)
IK 04 (per NF EN 50102 Ed. 95)
7.7
Variations in operating range
Influential
quantity
of influence
Battery voltage
9 V -12 V
Temperature
-10°C...+55°C
Humidity
10%...80% HR
Frequency
15...500 Hz
AC voltage
superimposed on
0% Un...20% Un
the test voltage
(1) The terms DAR, PI, DD and the capacity and current leak measurements are included in the
quantity "M Ω "
Range
Quantity
influenced (1)
V
MΩ
V
MΩ
V
MΩ (10 kΩ to 40 GΩ)
MΩ (40 GΩ to 10 TΩ)
V
MΩ
62
Influence
typical
Max.
< 1 pt
< 1 pt
0,15% read./10°C
0,3% reading +1 pt
0,20% reading
1% reading +1 pt
0,2% reading
1% reading +2 pt
0,2% reading
1% reading +5 pt
3% L
15% L +5 pt
0,3% reading
0,5% reading +1 pt
0,1% read./% Un
0,1% read./% Un +5 pt
2 pt
3 pt