SELECT
When the motor is not running, pressing the SELECT button will highlight the torque,
speed, or auto-reverse box. The function of the selected setting can now be adjusted.
When the motor is running, the SELECT button does not function.
2. Reading the Display
The selected box (torque, speed, or auto-reverse) is highlighted with a double-square.
In Figure 9, torque is selected.
TORQUE
• The torque limit values can be adjusted.
• Torque can be found in the top box.
• If the torque limit is exceeded, the file will stop and "S.t" is shown in the torque
box on the display.
Display
SPEED
• The rotation speed can be adjusted.
Display
Fig. 9
The present torque limit value is displayed. The
torque setting can be adjusted from 0.6 N
1.0 N
cm in increments of 0.2 N
•
1.0 N
cm to 4.0 N
•
The present rotation speed is displayed in the
middle box on the motor handpiece. The speed
setting can be adjusted from 100 to 800 rpm in
increments of 50 rpm.
cm, and from
•
cm in increments of 0.5 N
•
cm to
•
cm.
•
15