Technical Specifications - Microlife FR 1DM1 Manual Del Usuario

Termómetro de frente
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

13. Technical Specifications

Type:
Measuring Range: 34.0 °C to 42.2 °C (93.2 °F to 108.0 °F)
Resolution:
Accuracy:
Display:
Acoustic:
Memory:
Backlight:
Operating
temperature:
Storage/transport
temperature:
Automatic
Switch-off:
Battery:
Dimensions:
Weight:
Standards:
11
Forehead Thermometer FR 1DM1
0.1 °C / °F
Laboratory: ±0.2 °C, 36.0 ~ 39.0 °C
(±0.4 °F, 96.8 ~ 102.2 °F)
(Ambient Temperature: 16-40 °C)
Liquid Cristal Display, 4 digits plus special
icons
a. The unit is turned ON and ready for the
measurement: 1 short «bi» sound.
b. Complete the measurement: 1 long beep
(1 sec.), if the reading is less than 37.5 °C
(99.4 °F), 10 short «beep» sounds, if the
reading is equal to or greater than
37.5 °C (99.4 °F).
c. System error or malfunction: 3 short «bi»
sounds.
a. Auto-Display the last measured tempera-
ture when switched on.
b. 12 readings recall in the Memory Mode.
a. The display will be lighted GREEN for
4 seconds, when the unit is turned ON.
b. The display will be lighted GREEN for
5 seconds, when a measurement has
been completed with a reading less than
37.5 °C (99.4 °F).
c. The display will be lighted RED for
5 seconds, when a measurement is
completed with a reading equal to or
higher than 37.5 °C (99.4 °F).
16 °C to 40 °C (60.8 °F to 104 °F)
-20 °C to +50 °C (-4 °F to 122 °F)
Approx. 1 minute after last measurement
has been taken.
CR2032 BATTERY (X1) - at least 1000
measurements
122.36mm (L) x 42.2mm (W) x 43.11mm (H)
48g (with battery), 45g (w/o battery)
Complies with ASTM E-1965-98
requirements

Publicidad