Installation/Startup - Satelec Acteon Mini LED III SuperCharged Manual De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 34
MLED III Supercharged U7
Charging base:
Input voltage:
Protection:
Category:
Protection index:
Battery:
Type:
Dimensions:
Capacity:
Optical specifications:
- LED for polymerization:
Wavelength range:
Central wavelength:
Intensity:
Maximum exposure time:
Temperatures:
Operation:
Storage:
Humidity:
Operation:
Storage:
Atmospheric pressure
IV – INSTALLATION/STARTUP
4.1 UNPACKING THE UNIT
On reception of the unit, look for any damage that may
have occurred during transportation.
If necessary, contact your supplier.
4.2 RECOMMENDATIONS
Check that the environmental conditions have been
complied with (ambient temperature between 10°C and
40°C and humidity between 30% and 75%).
4.3 INSTALLATION
Ensure that the unit is not installed adjacent to
or on top of any other equipment.
Do not place the power cord in a cable feedthrough or
cable cover.
V6:Mise en page 1
05/09/2011
Remove the protective caps from the handpiece, screw
12 V DC
the battery onto the handpiece and insert the sterilized
3 AT F1 fuse (non
light guide into the handpiece.
accessible) 125 V
Ensure that the light guide is properly inserted,
Continuous operation
confirmed by a click.
IPX0
Ensure that all the segments of the LCD screen are present
once the battery has been properly screwed down.
Place the base on a hard and stable surface that is not
Lithium-Ion
inclined by more than 5°.
88 x Ø24 mm
Connect the AC adapter after ensuring that the voltage
2300 mAh
indicated corresponds to the electrical installation
available in the dental practice.
Two audible beeps confirm that the lamp has been
correctly placed on its holder.
420 - 480 nm
The translucent holder base lights up in blue and starts
455 – 465 nm
flashing, indicating that the battery is charging.
2000 mW/cm² ± 10% for
As soon as the battery is charged, the translucent holder
an active fiber diameter
base stops flashing and the blue light remains
of 7.5 mm
illuminated continuously.
15 seconds (show 15)
4.4 OPERATING FOR THE FIRST TIME
Before using for the first time:
+10°C to +40°C
- The delivered batteries are loaded at 60% before
-20°C to +70°C
- it is essential that the sterilizable accessories (light
30% to 75%
10% to 100%
condensation included
700hPa to 1060hPa
4.5 MODES AVAILABLE
The Mini LED SuperCharged has three modes you can
choose by pushing the left button (Fig.2-6) :
-
-
-
4.6 EXPOSURE TIMES AVAILABLE
In these modes, you can choose different exposure
times.
By pushing the right button (fig.2-7), you can select the
following combinations:
- Fast-Cure mode:
- 3 seconds (show 03) ;
- 4 seconds (show 04) ;
- 5 seconds (show 05) ;
- 10 seconds (show 10) ;
- Pulse-Cure mode:
- 5 shots of 1 second (full power with emission of 5
10
09:53
Page 10
shipping. It is therefore necessary to proceed to a
complete loading.
guide and protective shield) be sterilized and that the
unit base be disinfected (see chapter concerning
maintenance).
Fast-Cure,
Pulse-Cure,
Step-Cure.