MLED III Supercharged U7
The Mini LED SuperCharged must not be used near or on top of another device.
If this cannot be avoided, its operation under the conditions of use must be checked beforehand.
The use of accessories other than those specified or sold by SATELEC as replacement parts may have the consequence
of increasing the emissions or decreasing the immunity of the Mini LED SuperCharged.
10.1 - Electromagnetic emissions
The Mini LED SuperCharged is intended for use in the electromagnetic environment specified in the table below. The
user and/or installer must ensure that the Mini LED SuperCharged is used in such an environment.
Emission test
RF emission - CISPR 11.
10.2 - Magnetic and electromagnetic immunity
The Mini LED SuperCharged is intended for use in the electromagnetic environment specified in the table below.
The user and/or installer must ensure that the Mini LED SuperCharged is used in such an electromagnetic
environment.
Immunity test
Magnetic field at 50Hz.
3A/m
IEC61000-4-8
Electrostatic discharge
± 6KV contact
(ESD)
± 8KV air
IEC 61000-4-2.
Electrical fast transients
± 2KV for power supply
IEC 61000-4-4.
lines
Surges
± 1KV differential mode
IEC 61000-4-5.
± 2KV common mode
<5% UT
for 0.5 cycles.
Voltage dips, short
40% U
interruptions and voltage
5 cycles
variations
70% U
IEC 61000-4-11.
25 cycles
<5% U
250 cycles
10.3 - Electromagnetic immunity / mobile radiofrequency equipment
The Mini LED SuperCharged is intended for use in the electromagnetic environment specified in the table below.
The user and/or installer must ensure that the device is used in such an electromagnetic environment.
V6:Mise en page 1
Compliance
Mini LED SuperCharged uses RF energy for internal operation. Therefore, its radiofrequency emissions
are very low and are not likely to cause any interference in nearby equipment.
Group 1
Mini LED SuperCharged is suitable for use in all establishments, including domestic establishments and
Class B
those directly connected to the public low voltage power supply network that supplies buildings used for
domestic purposes.
IEC 60601 test level
Compliance level
3A/m
± 6KV contact
± 8KV air
± 2KV for power supply
lines
± 1KV differential mode
± 2KV common mode
(>95% dip in U
)
<5% U
(>95% dip in U
T
T
T
0.5 cycles.
(60% dip in U
) for
40% U
(60% dip in U
T
T
T
cycles
(30% dip in U
) for
70% U
(30% dip in U
T
T
T
25 cycles
(>95% dip in U
) for
<5% U
(>95% dip in U
T
T
T
250 cycles
05/09/2011
09:53
Electromagnetic environment - guidance
Electromagnetic environment - guidance
The intensity of the magnetic field should be equivalent to that of
a typical commercial or hospital environment (hospital, clinic).
Floors must be wood, concrete, cement or tiled.
If floors are covered with synthetic material (carpet, etc.), the
relative humidity must be at least 30%.
Power quality should be that of a typical commercial or hospital
environment.
Power quality should be that of a typical commercial or hospital
environment.
) for
T
Power quality should be that of a typical commercial or hospital
) for 5
T
environment.
If the use of the Mini LED SuperCharged requires continued
) for
T
operation during a power cut, it is recommended that the
product be powered from a separate power supply (UPS, etc.).