Manual de instrucciones del VisualEyes™ 515/525 - ES
Radiated RF
Immunity
IEC 61000-4-3
Electrical fast
transient/burst
IEC 61000-4-4
Surge immunity
IEC 61000-4-5
Conducted RF
Immunity
IEC 61000-4-6
Power Frequency
Magnetic Immunity
IEC 61000-4-8
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
NOTE: U
is the mains voltage prior to application of the test level.
T
Stress 4.5 V/M
+/- 2 kV for power supply
lines
+/- 1 kV for input/output
lines
+/- 1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
Stress 3 V and 10 V
30 A/m
0% U
10 ms
T
40% U
100 ms
T
70% U
500 ms
T
0% U
5000 ms
T
Stress 4.5 V/M
+/- 2 kV for power
supply lines.
No input/output lines
+/- 1 kV line(s) to line(s)
+/- 2 kV line(s) to earth
Stress 3 V and 10 V
30 A/m
0% U
10 ms
T
40% U
100 ms
T
70% U
500 ms
T
0% U
5000 ms
T
Página 40
Not Applicable
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Not applicable
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of the VNG
requires continued
operation during power
mains interruptions, it is
recommended that the
VNG be powered from
an uninterruptible power
supply or battery.