Intended Use; Queries And Technical Problems - Miele A 104 Instrucciones De Manejo

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 45
Re-usable medical devices can be reprocessed using this upper bas‐
ket in a Miele washer-disinfector. Follow the instructions for your
washer-disinfector as well as the instructions of the medical device
manufacturer on how to reprocess their items by machine.
Internal cleaning and disinfection to an appropriate standard is only
possible with instruments which can be sufficiently dismantled and
with adequate use of injector nozzles, irrigation sleeves and adapters.
The reprocessing result of the inside of instruments must be checked
as necessary using a test other than a visual check.
The A 104 upper injector basket is designed for the machine reproc‐
essing of surgical MIS, arthroscopy and urology instruments.
It is divided into two sections:
– The right-hand side is fitted with an injector manifold with 11 silico‐
ne holders for injector nozzles and irrigation sleeves for reproces‐
sing hollow instruments such as trocars, cannulae and the shafts of
MIS instruments.
– The left-hand side is free and can be equipped with inserts and
mesh trays for reprocessing instruments without lumen.
The water intake pipe also has several connections for hose adapters,
e.g. for E 448 hoses. Connectors not in use must be fitted with E 362
blind stoppers..
This upper basket can be used in the following washer-disinfectors:
– PG 8582
– PG 8582 CD
– PG 8592
Medical devices for reprocessing are referred to as the wash load if
they are not more closely defined.

Queries and technical problems

If you have any queries or technical problems that you would like to
discuss please contact Miele. Contact details can be found at the
back of the Operating Instructions for your Washer-Disinfector or go
to www.miele-professional.com.
en - Intended use
33

Publicidad

Tabla de contenido
loading

Tabla de contenido