Hach Flow Paddlewheel Sensor Manual Del Usuario página 11

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Operation
User navigation
Refer to the controller documentation for keypad description and
navigation information.
Configure the sensor
Use the Configure menu to enter identification information for the sensor
and to change options for data handling and storage.
1. Push the MENU key and select Sensor Setup, [Select Sensor],
Configure.
2. Use the arrow keys to select an option and push ENTER. To enter
numbers, characters or punctuation, push and hold the UP or DOWN
arrow keys. Push the RIGHT arrow key to advance to the next space.
Option
Description
SELECT CHANNEL Shows the channels to be configured.
MODULE NAME
Sets the name of the module.
RESET DEFAULTS Sets the configuration menu to the default settings. All
sensor information is lost.
CHANNEL A–D
Sets multiple options for the desired channel.
3. For each selected channel, use the arrow keys to select an option and
push ENTER.
Option
Description
SENSOR NAME
Changes the name that corresponds to the sensor on the
top of the measure screen. The name is limited to 4
characters in any combination of letters, numbers, spaces
or punctuation.
FLOW FORMAT
Sets the flow units, the multiplier and the resolution.
LOW FLOW
Sets the minimal level for liquid flow. If the flow goes below
CUTOFF
this level, a warning appears.
VOLUME UNITS
Sets the units for volume (e.g., liters).
Option
Description
TOTALIZER
Sets the totalizer volume to Auto (default) or Manual.
MODE
TOTALIZER
Resets the totalizer volume. Appears only if the totalizer
RESET
mode is set to Manual.
FILTER
Sets a time constant to increase signal stability. The time
constant calculates the average value during a specified
time—0 (no effect, default) to 60 seconds (average of
signal value for 60 seconds). The filter increases the time
for the sensor signal to respond to actual changes in the
process.
SELECT
Sets the tee-mount sensor type.
SENSOR
SET SENSOR
Sets the tee size and material. These settings are
dependent on each other.
LOG SETUP
Sets the time interval for data storage in the data log—5,
FLOW
30 seconds, 1, 2, 5, 10, 15 (default), 30, 60 minutes.
LOG SETUP
Sets the time interval for data storage in the data log—5,
VOLUME
30 seconds, 1, 2, 5, 10, 15 (default), 30, 60 minutes.
SERIAL NUMBER Sets the serial number of the sensor, limited to 16
characters in any combination of letters, numbers, spaces
or punctuation.
Calibrate the sensor
About sensor calibration
The sensor characteristics slowly shift over time and cause the sensor to
lose accuracy. The sensor must be calibrated regularly to maintain
accuracy.
During calibration, data is not sent to the datalog. Thus, the datalog can
have areas where the data is intermittent.
Calibration procedure
Note: If the controller is powered on when the paddle wheel sensor is not connected,
the calibration history is deleted.
English 11

Publicidad

Tabla de contenido
loading

Tabla de contenido