7) When F4 oFF is displayed press the [Print] key to return to weighing or press
[Mode] to go to the next function.
6.5
SELECTING THE INTERNAL OR EXTERNAL CALIBRATION
You can select if the internal mass or an external mass is to be used to calibrate the
balance.
1) To set this parameter press the [Mode] key during self-checking test when
turning on the balance.
2) After a few seconds, the display will show the first function F1 UNT.
3) Keep pressing the [Mode] key until F5 IEC is displayed.
4) Press the [Tare] key to view the current settings.
5) To change the settings press the [Mode] key to change from INt (internal
calibration) or E (external calibration)
6) Press [Tare] to store a particular setting. The display will return to F5 IEC
7) When F5 IEC is displayed press the [Print] key to return to weighing or press
[Mode] to go to the next function.
6.6
ADJUST THE VALUE OF THE INTERNAL MASS
The internal mass value stored in memory can be adjusted to more closely match
the value of the user's external mass.
1) To set this parameter press the [Mode] key during self-checking test when
turning on the balance.
2) After a few seconds, the display will show the first function F1 UNT.
3) Keep pressing the [Mode] key until F6 CA is displayed.
4) Press the [Tare] key to view the current settings.
5) The display will show the current value with the first digit flashing. To change
the value press [%] to shift the position, press [Mode] to increase the value
© Adam Equipment Company 2014
18