21. Specifications
Mechanical
• Dimensions:
• Weight:
Operational
• Flow Rate:
• Volume:
• Increments:
• Feed Until Empty (Continuous)
• Priming Rate:
•
Accuracy: ± 5% according to the following parameters:
Rate:
Volume:
Liquid:
Administration Set:
Up to 1500 mL delivered volume*
*Delivery of more than 1500 mL with one set may affect accuracy.
Head Height above rotor:
• Head Height Effects on Accuracy: ≤ 2.5% per 30cm of head height.
Head height range:
• Back Pressure Effects:
•
Occlusion Detection Pressure:
Upstream:
Downstream:
Note:
Low flow rates may significantly extend the time to occlusion alarm
• Air Detection
• Operating Temperature:
• Operating Humidity:
• Operating Atmospheric Pressure:
Storage
• Storage Temperature:
• Storage Humidity:
• Atmospheric Pressure:
L:140 mm x B:140 mm x H:62 mm.
499 g
1 mL/hr to 600 mL/hr
1 mL to 4000 mL
1 mL
CONT
≥700 mL/hr
100 mL/hr
100 mL
Water
New, previously unused set
30cm
-30cm to 60cm
≤ 0.5% per PSI
-34 kPa for 5 seconds
83 kPa for 30 seconds
103 kPa for 5 seconds
1.5 mL
+5° to +40° C
15% to 93%, non-condensing
65 to 106 kPa
-20°C to +60°C
10% to 95%, non-condensing
50 to 106 kPa
Power
• Voltage:
Input:
Output:
• Fuse Type:
• Battery Type:
• Battery Charge Time:
• Battery Life:
• Battery capacity decay due to:
Storage:
For optimal battery life, do not expose battery to elevated
storage temperatures for extended periods. Store below
50°C when possible and do not exceed 60°C storage
temperature. For extended storage, store battery at low
temperature and low charge state.
Aging:
Please note that the battery capacity numbers noted below
are estimates and that actual battery capacity may vary sig-
nificantly depending on factors such as actual use, charge
cycles and temperature.
Pump Age
Approx Run Time at 125 mL/Hr
1 yr
24 hrs
2 yr
24 hrs
3 yr
20 hrs
4 yr
16 hrs
5 yr
12 hrs
Above 5 years
Service Pump*
* Service interval for the Compat Ella
will be replaced.
In the case of a malfunction, contact your Nestlé Health Science repre-
sentative or Nestlé approved service provider to arrange for the pump to
be repaired or replaced.
• Electric shock protection:
• Mechanical and Fluid ingress:
The Compat Ella
is designed in accordance with EN 60601-1-2 ed.3.
®
standard for electromagnetic emissions and immunity.
EN
100-240 V; 50-60 Hz;
5 VDC 2A
Single blow fast acting
Lithium Ion
≤ 6 hours
24 hours, at 125mL/hr
pump is every 5 years. Battery
®
Class II
IP25
61