4 OPERATION
2.
Press the Max Inflate button again to manually disengage Max Inflate mode.
This will deactivate the Max Inflate LED.
To protect the patient from lying on a firm surface too long, the pressure will automatically return to normal range after 20 minutes.

Lock Button

1.
Press the Lock button (
Press the button for approximately 3-5 seconds for the Lock LED to activate.
The Lock LED illuminates when in this mode.
Alarm Silence Button
1.
Press the Alarm Silence button (
the control unit.
The Alarm LED illuminates when in this mode.
Power Fail LED
In the event of power outage, an alarm sounds and Power Fail LED flashes AMBER. The control unit has internal memory and retains the previous
settings during the power outage.
During a power outage, the overlay retains the air as long as the overlay is connected to the control unit.
Low Pressure LED
In the event that the overlay hose disconnects, an alarm sounds and Low Pressure LED flashes AMBER. Once the low pressure problem is fixed, the
control unit resumes operation in the previously set mode.
Invacare® MA50CE Series
) to lock out all control unit functions, including the On button, to prevent any tampering with the settings.
) to silence the alarm that sounds in the event of power failure or when the hose is disconnected from
22
Part No 1176084