Mettler Toledo M80SM Manual De Instrucciones página 24

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 27
en
24
If configuration of an M80 SM transmitter is
necessary use the METTLER TOLEDO M80 Trans-
mitter Configuration Tool (TCT) software and the
METTLER TOLEDO Transmitter Configuration Box
(see chapter 9 for ordering information). Please
download the TCT software from the documenta-
tion section on www.mt.com/M80.
The M80 SM stores installation point-specific
MODBUS parameters. So there's no need to re-
configure the transmitter when a 1-wire sensor
is replaced.
5
Technical data M80 SM
Measurement parameters 1-wire sensors: pH,
ISM
Power supply
Operating temperature
Relative humidity
Protection class
6
Connections
The M80 SM 1-wire has one connector to fit any
METTLER TOLEDO 1-wire ISM sensor (AK9
connector).
h
M80 SM
30 530 859 A
NOTE: Important operating information
The M80 SM transmitter cannot be auto-
claved! Disconnect the M80 SM trans-
mitter from the sensor before autoclaving
the sensor!
M80 SM Sensor Mount Transmitter
amperometric Dissolved
Oxygen, Dissolved CO
Conductivity
Full ISM functionality
24 VDC (min. 100 mA) /
8 – 30 VDC (min. 2 W)
–15 to +60 °C
5 ... 95 % rH
(non-condensing)
IP65
© 08 / 19 Mettler-Toledo GmbH
Printed in Switzerland
,
2

Publicidad

Tabla de contenido
loading

Tabla de contenido