6.
Cleaning the unit / Maintenance
Caution
Cleaning the unit / Maintenance
•
Always turn off the unit and disconnect the mains cable from the power source
before cleaning the unit.
•
Prevent humidity from entering into the circulator.
•
Electrical connections and any other work must be performed by qualified
personnel only.
The Flow-Through Cooler is designed for continuous operation under
normal conditions. Periodic maintenance is not required.
Regularly check the condensor for dirt contamination. Clean the
ribbed condensor, because dust and dirt will reduce cooling
performance of the unit.
Cleaning the Cooling Compressor:
•
Switch off the unit, disconnect mains power cable.
•
Remove the hood.
•
Clean the ribbed condensor with a vacuum cleaner.
•
Replace the hood or the ventilation grid.
•
Switch on the unit.
Clean the outside of the unit using a wet cloth and low surface
tension water.
Repairs:
Before asking for a service technician or returning a circulator for
repair, please contact our service station.