5. INSTALLATION

5.1 SETTING UP THE LIGHTSOURCE
Place the illuminator on a stable surface (cart, counter, stand, etc.).
NOTE
Avoid places where the illuminator may be splashed with liquid.
Absolutely DO NOT use in any environment with explosive or
flammable gases.
This illuminator pulls cooling air into its cabinet through a filtered grille
in the bottom panel underneath the front portion and exhausts warm air
from the fan in the rear.
DO NOT block the space in front or below the front of the illuminator or
impede exhaust air flow behind or above the cabinet.
A clear space minimum of 5"/12.7cm behind and above the cabinet is
required.
The illuminator should not be placed where its exhaust will influence
other devices, or where the heated air from other devices will influence
the illuminator.
Ensure the rear power switch is in OFF position.
Ensure that the lamp is properly positioned for use. Open the lamp compartment door.
Move the lamp positioning lever (found to the right of the lamp) clockwise, from
horizontal to vertical. As the positioning cones disengage the lamp should drop about
3mm. Move the lamp positioning lever counterclockwise to reengage the positioning
cones and observe that the lamp rises about 3mm as it returns to operating position.
Close the lamp compartment door.
Connect AC power cord to the power inlet located on the rear panel of the illuminator.
Ensure the power cord is completely seated into the power inlet and engage the cord
retainer if so equipped.
Use only power cords provided with the illuminator.
CAUTION
Plug the AC power cord into a wall outlet using the power cord plug supplied with the unit.
To prevent electric shock, connect power cords of peripheral equipment
through medical isolation transformers.
CAUTION
NOTE
When using medical isolation transformer, be sure to check the
power ratings.
LIT-217 Sunoptic Surgical
Rev. D
®
Page 9 of 112
transformer
(English)