This illuminator is designed for exclusive use with the ST TECHNOLOGIES p/n
SYC0300 lamp module. Failure to use this lamp could void the warranty.
Open the lamp compartment door.
Move the lamp positioning lever (found to the right of the lamp) clockwise, from horizontal to
vertical, and pull the lamp cartridge out.
Grasp only the horizontal top and bottom grip flanges of the plastic housing of the lamp, pull
straight out to remove the lamp. Insert a new one (lamp pointed to the left), pushing firmly
to assure full engagement to the power connectors. Lower the lamp engagement lever
counter-clockwise, from vertical to horizontal, and observe that the lamp rises about 3mm
as it moves to its proper operating position.
Log the serial number and hours on the hour meter. Close the lamp compartment door.
Re-connect the power cord and turn the light source on according to 6.1. Reset hour meter
to zero by depressing hour meter reset button on rear panel.
If the tamper seal on the lightsource is broken, the WARRANTY WILL BE
NOTE
VOIDED. There are no other user serviceable parts in the unit.
8.2 LIMITED WARRANTY
Your illumination product(s) carry(s) a 3-year warranty from the date of shipment on
workmanship and all defects of material, excluding, fiberoptic cable and lamps. Should your
product prove to have such defects within three years of the shipment, ST Technologies
will repair or replace the product or component part without charge.
illumination product(s) need servicing under this warranty, please contact ST Technologies
®
for return authorization documentation. Please carefully pack the unit in a sturdy carton
and ship it to the factory. Include a note describing the defects, your name, telephone
number and a return address.
accidental damage, normal wear and tear or if transferred to a new owner without
authorization from ST Technologies
you may also have other rights that vary from state to state.
POST WARRANTY REPAIRS: You may return your product(s) for repair, shipping prepaid
to the factory. Your product will be inspected and an estimate of repair charges will be
submitted to you for approval.
•
In the US call:
•
FAX number:
®
LIT-217 CUDA
SURGICAL
(English)
Warranty does not cover equipment subject to misuse,
®
. This warranty gives you specific legal rights and
877 677-2832
904 733-4832
Should your
(toll free)
Rev. B
Date of Revision: 05/02/17
Page 12 of 71
®