Descargar Imprimir esta página

Storage And Maintenance; Troubleshooting Guide - bewell connect MyTens Pro Manual Del Usuario

Ocultar thumbs Ver también para MyTens Pro:

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

2. Stopping a session : Click on the Stop Button.
3. Estimate your satisfaction level (the satisfaction level is available in TENS mode only). The ses-
sion is then recorded in the history.
4. To turn off the device, press the "On/Off" button.
5. Remove the electrodes from the skin. Detach from the edges.
6. Remove the device from the electrodes. If you forget the device for more than 5 minutes, it
turns off.
If the device is frequently used during the day, it is advisable to charge it before starting a new
session.

6. STORAGE AND MAINTENANCE

6.1. TORAGE OF THE ELECTRODES
Put the electrodes back into their protective bag. To prevent buildup of dust on the electrodes,
close the bag carefully. Prefer to store electrodes in a place where the temperatures is 5 to 27°C.
Avoid extreme heat and exposure to direct sunlight. Do not keep the electrodes in the freezer
or the refrigerator.
6.2. NETTOYAGE
Warning : Never immerse the product in water and do not rinse it. Never immerse the elec-
trodes in water. Do not use cleaning agents other than those mentioned below. They could
cause serious damage to the equipment.
1. Using a damp cloth and mild detergent (eg: dishwashing liquid), clean the device. It is also
possible to use isopropyl alcohol at 70° (IPA).
48
BewellConnect - BW-TS1 - User Manual - 072015
2. If dirt gets on the electrodes, put a drop of water on your finger and gently remove the dust
from the surface. Do not use soap or alcohol to clean the electrodes.

7. TROUBLESHOOTING GUIDE

Problem
The battery indicator on the
Discharged battery
device is displayed in orange
continuously
Orange battery light and then
Discharged battery
the device switches off.
Treatment which seems to be
1. Electrodes not placed in the
different from or less pleasant
correct location
than the previous one
2. Intensity too low or too high
3. Expiry date on the electrodes
exceeded
4. Worn electrodes
5. Electrodes too dirty
6. You did not correctly apply the
electrodes to the skin
Unpleasant feeling when touch-
The session has not been
ing the connectors of the device
stopped in the application or
or the electrodes
the on/off button has not been
pressed before removing the
device
8. AVOIDING MALFUNCTIONS
- Do not perform stimulations in the vicinity of electronic surveillance devices (such as cardiac
monitors or electrocardiogram alarms). They might not work properly when using the electri-
cal stimulation device at the same time.
- Do not operate the device at less that 1 metre from shortwave or microwave medical equip-
ment. Close proximity with this device may cause a degradation of performance of the device.
- Portable RF communications equipment (including devices like antenna cables and external
antennas) should not be used at less than 30 cm from the least part of the device, including
cables mentioned by the manufacturer. This could lead to deterioration of performance of the
device.
BewellConnect - BW-TS1 - User Manual - 072015
Possible cause
Solution
Recharge the device
Recharge the device
1. Switch off the device and
remove the electrodes. Remove
the electrodes and move them
slightly.
2. In the App, modify the inten-
sity level.
3. Remove the electrodes and
replace them with new ones.
4. Replace the electrodes.
5. Clean the electrodes
(paragraph 6.2 "Cleaning"). If the
problem occurs again, replace
the electrodes.
6. Switch off the device and
remove the electrodes. Make
sure the electrodes are properly
placed on the skin.
Always end the session in the
application or remember to
press the on/off button before
removing the device.
49

Publicidad

loading

Productos relacionados para bewell connect MyTens Pro

Este manual también es adecuado para:

Bw-ts1