Administrative Mode - Medela Invia Liberty Manual De Uso

Ocultar thumbs Ver también para Invia Liberty:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Administrative Mode

The default settings of the Invia Liberty Pump are -60 mmHg and constant mode.
CAUTION
The pressure level should always be set according to Physician's instruction.
The set pressure is measured and controlled at the end of the conocal connector
on the canister tube.
1. Switch Invia Liberty Pump ON in Administrative mode,
press and hold [
Self-test starts
If the self-test is not successful, follow instructions shown on
the display or see the Alarm chapter.
2. Acknowledge disclaimer. Press
"New patient? Yes / No"
"Yes"
means that Invia Liberty Pump will issue a new
therapy ID number (= new patient).
This number can be noted in the patient's file.
The therapy ID number is displayed in the information field.
"No"
means that the therapy ID number and settings
remains unchanged (= same patient).
], press [
].
[
"OK"
] to confirm.
21